Baptiste Hériard-Dubreuil, Adrien Besson, Jonathan Mamou, Joël Gay, Juliette Foucher, Victor De Lédinghen, Claude Cohen-Bacrie
Ultrasound in Medicine & Biology, September 2024
Abstract
Objective
This study aimed to evaluate the performances of quantitative ultrasound (QUS) for the detection and assessment of hepatic steatosis when implemented using an ultraportable ultrasound scanner.
Methods
Seven established QUS parameters were investigated. Ultrasound signals were acquired using a new ultraportable ultrasound device, the Hepatoscope. The feasibility of QUS using the Hepatoscope was first assessed in vitro. Clinical reliability, accuracy and staging capabilities were evaluated in 60 patients referred to a hepatology consultation for known chronic liver disease and enrolled in a prospective clinical investigation using the controlled attenuation parameter (CAP) as ground truth.
Results
QUS parameters showed moderate (intra-class correlation coefficient [ICC] >0.50) to excellent (ICC >0.90) reliability. Two parameters, namely Lizzi-Feleppa mid-band fit and attenuation, were both reliable (ICC = 0.89 and 0.86, respectively) and correlated with the CAP (squared Pearson correlation coefficient of R2 = 0.65 and R2 = 0.6, respectively). For steatosis detection (S0 vs. ≥S1), the two parameters yielded an area under the receiving operating characteristic curve of 0.90 and 0.86, respectively (95% confidence interval: [0.81–0.99] and [0.76–0.96], respectively).
Conclusion
QUS can be reliably and accurately implemented on ultraportable ultrasound scanners. The combination of ultraportability and quantitative assessment of liver fat is promising for large-scale screening and monitoring of hepatic steatosis.